    Ova Plus Electrodes 4 Packs x 2 pads

    There are THREE positions where the pads can be placed for maximum benefit. It is recommended that the positions are tried in the following order until maximum relief is obtained: • On the lower back • On the front, in a vertical format • On the front, in a diagonal format
    Product #: 26483T

    Please refer to the diagrams below:

    On the lower back:

    Place the pads vertically either side of the
    spine, approximately 10cms (4 inches) apart,
    with the bottom of the electrodes no higher
    than the top of your buttocks. The pigtails
    should hang down from the pads.

    In order to obtain the best relief, the pads need to be placed over the dermatomes T10, T11 and T12, which are the nerve roots that supply the sensory fibres to the uterus. These dermatomes wrap around the body, but do not quite meet up at the front. They run from around waist height at the back down to the groin area at the front. As each person’s body is different, the Ova+ is supplied with pads which are large enough to be effective even if not located exactly over all three dermatomes. When the pads are placed correctly the Ova+ will work very well so if after thirty minutes considerable relief has not been achieved, either move the pads slightly or try one of the other pad positioning areas.

    The self-adhesive pads supplied are intended for multi-use by one person only. Do not lend used pads to others, nor use someone else’s used pads yourself.

    To apply the pads simply peel off the clear film liner and the pads will adhere to the skin automatically. Return the plastic film liner into the bag, as the pads will need to be replaced onto the film after use.

    Points to note when applying the pads:

    • Make sure the Ova+ is switched off before applying or removing the pads.

    • Ensure that the skin is clean and thoroughly dry before attaching the pads, as they will not adhere well if the skin is wet or greasy.

    • Do not worry if a small part of each pad covers the pubic hair. After use the pads can be removed gently without pulling out the hair.

    • If a large part of the pad area is covered with hair the stimulation may be less effective. If hair needs to be removed, wait 48 hours after shaving, waxing or using depilatory cream before applying the pads, to avoid irritating the skin.

    • Remove the pads from the skin or from the liner by peeling them back from the edge of the pad. Do not remove by pulling on the pad’s pigtail wire.

    • If you find that a small part of the pad overlaps the top of your thigh, and that this is uncomfortable, then it is possible to cut off this part using scissors. Before cutting make sure that you are cutting from the opposite end to the lead wire and that you are not cutting through any of the wires which run through the pad. Do not cut off more than a small triangular piece, with sides no more than 20mm long and an area no greater than 200 sq.mm. See diagram. Ensure that the leadwire is slack at all times when the Ova+ is in use. Excessive force applied to the leadwire could cause a pigtail to separate from its pad.

    Care of Electrode Pads

    The electrodes supplied with your Ova+ unit are self-adhesive and can be used several times. Skin must be allowed to breathe, so the pads should be removed periodically. When not in use, the pads should be placed onto the plastic film liner and sealed in the bag.

    • When the pads initially lose their adhesive quality, it is possible to reactivate their adhesiveness by applying a fine mist of water.

    • Replace pads when they lose their stickiness. Poor connection may cause discomfort and skin irritation.

    • Storage life of an unopened pack of electrodes is 2 years. This may be affected by very high temperatures or very low humidity.

    • The pads provided are latex-free.